Skip to content
Retina & Vitreous

Electrooculogram (EOG)

An electrooculogram (EOG) is an electrophysiological test that records the standing potential present in the eye using skin electrodes placed at the outer canthus. This standing potential is positive at the cornea and negative at the posterior pole (Bruch’s membrane side), with a potential difference of about 6 mV in a healthy eye. The signal amplitude recorded during an actual test is typically around 250 to 1,000 μV.

The standing potential of the EOG indirectly reflects the transepithelial potential (TEP) of the RPE. This potential difference, which changes in response to light stimulation, is recorded over time, and the ratio of the dark trough to the light peak is calculated to evaluate RPE function.

The EOG was described and named by Erwin Marg in 1951. In 1962, Geoffrey Arden reported the clinical usefulness of the Arden ratio, leading to its widespread use as a diagnostic test for fundus diseases. Currently, the International Society for Clinical Electrophysiology of Vision (ISCEV) sets the standard, with the latest version published in 2017.

Unlike the electroretinogram (ERG), which evaluates the function of photoreceptors and bipolar cells, the EOG primarily reflects the functional integrity of the RPE. Therefore, it is useful for diagnosing diseases where the ERG is normal but the EOG is selectively abnormal. The test takes about one hour, so its frequency in general clinical practice is limited.

Q What is the difference between EOG and ERG?
A

The ERG mainly evaluates the function of the neural retina, including photoreceptors (cones and rods) and bipolar cells. The EOG reflects the functional integrity of the RPE (retinal pigment epithelium). In Best vitelliform macular dystrophy, the ERG is normal but the EOG is abnormal, making the combination of these two tests useful for diagnosis.

Subjective Symptoms (Common Complaints in Diseases with Abnormal EOG)

Section titled “Subjective Symptoms (Common Complaints in Diseases with Abnormal EOG)”

The EOG itself has no subjective symptoms. The following symptoms are observed in retinal and RPE diseases that show abnormal EOG.

  • Decreased visual acuity: Central vision loss due to macular RPE damage.
  • Metamorphopsia: Objects appear distorted. Common in macular diseases.
  • Impaired dark adaptation and night blindness: Occurs in diseases involving rod dysfunction.
  • Central scotoma: Visual field defect around the fixation point. Occurs in macular dystrophy.

Clinical Findings (Interpretation of EOG Values)

Section titled “Clinical Findings (Interpretation of EOG Values)”

The Arden ratio (L/D ratio: light peak ÷ dark trough) obtained from the EOG test is the main evaluation index.

The criteria for the Arden ratio are shown below.

JudgmentArden ratioClinical significance
Normal≥1.80Normal RPE function
Borderline1.65–1.80Further examination required
Abnormal<1.65Widespread RPE dysfunction

A value less than 1.5 strongly suggests widespread outer retinal damage. The 2017 ISCEV recommends the term “light peak to dark trough ratio” over Arden ratio.

Characteristic findings of the EOG waveform are as follows:

  • Dark trough: The potential reaches its minimum after 10–15 minutes of dark adaptation. It is a non-photosensitive component and reflects the structural integrity of the RPE itself.
  • Light peak: The potential reaches its maximum value 7–12 minutes after light adaptation. It is a light-sensitive component and reflects activation of the RPE basal membrane.

Supernormal response in MEWDS (multiple evanescent white dot syndrome)

Section titled “Supernormal response in MEWDS (multiple evanescent white dot syndrome)”

In MEWDS, a “supernormal response” has been reported in which the Arden ratio of the affected eye exceeds that of the fellow eye.

Wang F et al. (2024) evaluated a case of MEWDS by combining EOG with en-face IS/OS-ellipsoid zone (EZ) complex imaging. The Arden ratio of the affected eye (right eye) was 2.5, showing a supernormal response exceeding the fellow eye (left eye) ratio of 1.7. The dark trough was 5.0 min/422.0 μV in the right eye and 7.0 min/351.5 μV in the left eye, and the light peak was 19.0 min/1,051.1 μV in the right eye and 21.0 min/611.7 μV in the left eye1).

The mechanism of this supernormal response is unknown, but it is speculated to involve RPE hyperactivation during the acute inflammatory phase1).

Q What diseases are suspected when the Arden ratio is low?
A

Best vitelliform macular dystrophy is the most representative condition, where the EOG is selectively reduced even when the electroretinogram is normal. Other conditions such as fundus albipunctatus, choroideremia, chloroquine/hydroxychloroquine toxicity, and diabetic retinopathy (advanced cases) also show low values. For details, see the section “Diseases and conditions with abnormal EOG”.

3. Diseases and conditions with abnormal EOG

Section titled “3. Diseases and conditions with abnormal EOG”

Abnormalities in the EOG (decreased or normal Arden ratio) reflect the functional state of the RPE. Understanding the EOG findings for each disease is useful for diagnosis.

The EOG findings for major diseases are summarized below.

Disease nameEOG findingsNotes
Best diseaseMarkedly reducedElectroretinogram normal
Retinitis punctata albescensReduced to normalNo light rise after brief dark adaptation
Stargardt disease (advanced stage)DecreasedMay be normal in early stages
ChoroideremiaDecreasedWorsens with disease stage
Retinitis pigmentosaDecreasedRod-cone dystrophy similarly
Chloroquine toxicityDecreasedPersists after drug discontinuation

Major diseases showing decreased Arden ratio

Section titled “Major diseases showing decreased Arden ratio”
  • Best vitelliform macular dystrophy: An autosomal dominant disease caused by mutations in the BEST1 gene (bestrophin 1). The pattern of normal electroretinogram with isolated EOG reduction is specific and most useful for diagnosis.
  • Autosomal recessive bestrophinopathy (ARB): Caused by autosomal recessive mutations in the BEST1 gene. Fundus findings are diverse, and EOG is key for diagnosis.
  • Stargardt macular dystrophy (advanced stage): Macular dystrophy due to ABCA4 gene mutations. EOG may be normal in early stages.
  • Fundus albipunctatus: Caused by RDH5 gene mutations. No light rise is observed after 15 minutes of dark adaptation.
  • Choroideremia: Progressive atrophy of the RPE and choroid leads to decreased EOG.
  • Retinitis pigmentosa and rod-cone dystrophy: EOG is reduced in advanced cases with extensive RPE damage.
  • Gyrate atrophy of the choroid and retina: EOG is reduced due to RPE damage caused by ornithine metabolism disorder.
  • Chloroquine and hydroxychloroquine toxicity: RPE toxicity due to antimalarial drugs. May not improve after drug discontinuation.
  • Diabetes mellitus: EOG worsens with longer disease duration.
  • Intraocular foreign body (siderosis bulbi): EOG is reduced due to RPE damage from iron ions.
  • Choroidal malignant melanoma: May reflect RPE damage caused by the tumor.

In the following diseases, RPE function is preserved, so EOG is within the normal range.

  • Dominantly inherited drusen of Bruch’s membrane
  • Congenital achromatopsia (cone dysfunction, but RPE is normal)
  • Congenital stationary night blindness (rod dysfunction, but RPE is normal)
  • Optic nerve diseases (because the disorder is preretinal, EOG is not affected)

The following drugs are known to alter the EOG resting potential.

  • 20% mannitol intravenous injection: Reduces the resting potential by approximately 43%.
  • 500 mg acetazolamide intravenous injection: Lowers the resting potential.
  • Timolol: Affects the resting potential.

The standard EOG testing procedure follows ISCEV (2017 edition).

The main steps of the test are shown below.

StepTimeContent
Pre-adaptationAt least 15 minutesUnder 35-70 lux room lighting
Dark adaptation recording15-20 minutesDark room, red LED tracking
Light adaptation recording15–20 minutesGanzfeld illumination / LED tracking
  • Mydriasis is recommended.
  • Stay under stable room lighting (35–70 lux) for 30 minutes before the examination. At least 15 minutes of pre-adaptation is required.
  • Do not perform tests that use strong retinal illumination, such as fluorescein angiography (FA) or fundus photography, before the examination.
  • Attach silver-silver chloride disk electrodes to the skin near the inner and outer canthi.
  • Place the reference electrode on the center of the forehead or on the earlobe.
  • Place the ground electrode on the forehead.
  • Before attaching the electrodes, wipe off skin sebum with an alcohol swab to reduce impedance.
  • Connect the right electrode to positive (+) and the left electrode to negative (-).

While providing uniform light stimulation with a Ganzfeld dome, have the patient alternately track a red LED every minute (10 times per round trip).

  • Dark adaptation recording: Record every minute for 15–20 minutes in a dark room.
  • Light adaptation recording: Immediately after, record light adaptation for 15–20 minutes.
  • Plot the average amplitude values per minute to draw the EOG curve.

Noise Countermeasures and Test Limitations

Section titled “Noise Countermeasures and Test Limitations”
  • AC power noise, electromyography (EMG) artifacts, and electrical instability due to sweat are the main noise sources.
  • Apply electrodes after sweat has sufficiently dried.
  • In infants/elderly and patients with eye movement disorders (nystagmus, ophthalmoplegia), accurate tracking is difficult, and testing may be challenging.

Mandatory Reporting Items in the 2017 ISCEV Standard

Section titled “Mandatory Reporting Items in the 2017 ISCEV Standard”
  • Light peak to dark trough ratio
  • Dark trough amplitude (mV)
  • Time from light onset to light peak (minutes)

ISCEV Reporting Items

Light peak: dark trough ratio: Equivalent to Arden ratio. Normal is 1.80 or higher.

Dark trough amplitude: Absolute value of dark minimum (mV). Reflects RPE structural integrity.

Light peak time: Time elapsed from light onset (minutes). Usually 7–12 minutes.

Precautions during examination

Thorough pre-adaptation: 35–70 lux for at least 15 minutes. Avoid strong light stimuli.

Accuracy of pursuit: 5 round trips per minute. Difficult to perform if there is an eye movement disorder.

Beware of false normalization: If the baseline potential is extremely low, the L/D ratio may become falsely normalized.

ISCEV offers an optional test called “Fast Oscillations (FO).” It alternates dark and light phases every minute and reflects the function of the CFTR (cystic fibrosis transmembrane conductance regulator) chloride ion channel in the RPE basal membrane. It has been suggested that FO may be reduced in cystic fibrosis (CF).

Q How long does the EOG test take?
A

In addition to pre-adaptation (at least 15 minutes at 35–70 lux), dark adaptation recording takes 15–20 minutes and light adaptation recording takes 15–20 minutes, so the entire procedure takes about 1 hour. It may be difficult to perform in infants, elderly individuals, and patients with eye movement disorders because accurate tracking is challenging.

6. Pathophysiology and Detailed Mechanisms

Section titled “6. Pathophysiology and Detailed Mechanisms”

Mechanism of Generation of Resting Potential (TEP)

Section titled “Mechanism of Generation of Resting Potential (TEP)”

The resting potential recorded by EOG reflects the transepithelial potential (TEP) of the RPE. TEP is generated by the difference in membrane potential between the apical membrane and the basolateral membrane of RPE cells.

During dark adaptation, ion transport from photoreceptors changes, reducing ion influx into the RPE. As a result, the transepithelial potential of the RPE decreases, forming the dark trough. The dark trough is a non-photosensitive component and depends on the structural integrity of the RPE itself (cell density and cell membrane integrity).

During light adaptation, the following sequence of mechanisms causes RPE depolarization, increasing the potential and forming the light peak.

  1. Calcium (Ca²⁺) release from the endoplasmic reticulum in response to light stimulation
  2. Activation of bestrophin (BEST1 gene product) and L-type calcium channels
  3. Opening of calcium-dependent chloride (Cl⁻) channels
  4. Chloride ion efflux from the RPE
  5. RPE depolarization → TEP increase → light peak

The central role of bestrophin in this cascade explains why EOG is selectively abnormal in Best disease. In Best disease, BEST1 gene mutations impair bestrophin function, making it difficult to generate a light peak, thus reducing the Arden ratio.

  • Non-light-sensitive component (dark trough): Depends on the structural integrity of the RPE. It decreases when RPE cells are lost or degenerate.
  • Light-sensitive component (light rise): Depends on the depolarization mechanism of the RPE basolateral membrane. It depends on the function of bestrophin and calcium channels.
Q Why does the light peak occur?
A

Light stimulation induces Ca²⁺ release from the endoplasmic reticulum, opening calcium-dependent Cl⁻ channels mediated by bestrophin (the BEST1 gene product). Chloride ions are expelled from the RPE, causing RPE depolarization, which increases the transepithelial potential and forms the light peak. The selective reduction of EOG in Best disease is due to this bestrophin dysfunction.


7. Latest Research and Future Perspectives (Research-stage Reports)

Section titled “7. Latest Research and Future Perspectives (Research-stage Reports)”

Research on BCI/HCI technology that reads intentions from eye movements using EOG electrical signals has increased rapidly since the 2000s.

Belkhiria et al. (2022) reviewed the literature on EOG-based HCI from 2000 to 2020 and reported that applications for communication support for people with disabilities, wheelchair control via eye movements, and eye tracking are rapidly expanding 3).

EOG sensors have been integrated into eyeglass-type wearable devices such as J!NS MEME, enabling applications for monitoring eye movements, drowsiness, and concentration in daily life 3). Methods using infrared CCD cameras for direct recording of eye movements are also becoming widespread, gradually replacing conventional ENG (electronystagmography).

EOG (ENG) as a Recording Method for Eye Movements

Section titled “EOG (ENG) as a Recording Method for Eye Movements”

Electronystagmography (ENG), which applies the principle of EOG, is used for quantitative recording of eye movements. EOG is also standardly used as an eye movement channel in polysomnography (PSG).

Shoukat et al. (2022) reported PSG findings in a case of convergence-retraction nystagmus after midbrain hemorrhage. During wakefulness, nystagmus with a frequency of 2.8 Hz and amplitude of 60 μV was recorded on EOG 2). Nystagmus due to central nervous system (CNS) disorders usually disappears during sleep, but in this case, nystagmus persisted during both non-REM and REM sleep, which was considered characteristic of midbrain hemorrhage 2).

Significance of supernormal response in MEWDS

Section titled “Significance of supernormal response in MEWDS”

The mechanism by which a supernormal response, where the Arden ratio in the affected eye exceeds that in the healthy eye, occurs during the acute phase of MEWDS remains unknown. Acute inflammatory hyperactivation of the RPE may be involved, but elucidating the molecular mechanism is a future challenge 1).


  1. Wang F, Wang A, Leng X, et al. EOG and the En-Face Inner Segment/Outer Segment-Ellipsoid Complex Image in Multiple Evanescent White Dot Syndrome. Int Med Case Rep J. 2024;17:597-602.
  2. Shoukat U, Glick DR, Chaturvedi S, et al. Images: Polysomnographic findings of nystagmus caused by a midbrain hemorrhagic stroke. J Clin Sleep Med. 2022;18(5):1479-1482.
  3. Belkhiria C, Boudir A, Hurter C, et al. EOG-Based Human-Computer Interface: 2000-2020 Review. Sensors. 2022;22(13):4914.

Copy the article text and paste it into your preferred AI assistant.